Ludlum Model 3-90 Survey Meter

Radiation Safety Officer

 

 

Ludlum Model 3-90 Survey Meter

Specifications

Indicated Use – Alpha, beta, gamma survey
Detector – Internal pancake style G-M detector with 1.7 ± 0.3 mg/cm² window
Area – Active : 15.4 cm²
– Open : 12.2 cm²
Efficiency (2π geometry) 14C : 10%
90Sr / 90Y : 45%
99Tc : 38%
32P : 65%
239Pu : 30%
Sensitivity – Typically 3300 cpm/mR/hr (137Cs gamma)
Audio – Built in unimorph speaker with ON/OFF switch (greater than 60dB at 2 feet)
Meter Dial – 0 - 5 cpm, BAT TEST (others available)
Multipliers – X0.1, X1, X10, X100
Linearity – Reading within ±10% of true value
Calibration Controls – Accessible from front of instrument (protective cover provided)
Response – Toggle switch for FAST (4 seconds) or SLOW (22 seconds) from 10% to 90% of final reading
Reset – Pushbutton to zero meter
Power – 2 each "D" cell batteries (housed in sealed compartment that is externally accessible)
Battery Life – Typically 600 hours with alkaline batteries
Battery Dependance – Less than 3% change in readings to battery endpoint

 